4
Notes
1. It must be considered that this bioreactor can be configured to
operate in different modes of action, including compression,
tension and fluid flow. For each mode, the components used in
the assembly of the equipment’s chambers are different. In the
described method, the association of the compression modes
and fluid flow was chosen. The total number of mounted
chambers are suggested when sample comparison is a desired
parameter to be evaluated.
2. Medium characteristics may vary due to cell type.
3. The hermetic feature should be previously tested before auto-
claving through a manual sphygmomanometer.
4. The open/close apertures setup must allow the fluid flow
oriented as desired, that is, from top to bottom.
5. 3D porous scaffolds of materials sensitive to the autoclave
temperature can be disinfected with alcohol baths. To facilitate
the alcohol flow within the scaffold, the material can be
immersed in alcohol and subjected to vacuum. Successive
washes in saline solution, also subjected to the same vacuum,
will be necessary for the removal of alcohol.
6. Cell attachment and matrix synthesis during the preculture
period are important for a more effective mechanotransduction
and resultant tissue stiffening when under the dynamic-
mechanical environment [20]. One must also observe that
differentiated cell types must have distinct conduct as com-
pared to undifferentiated ones. Mature cells that exhibit high
matrix expression should be treated with additional care to
properly adapt the medium content.
7. According to the desired flow direction, one must choose
different inputs and outputs apertures, and previous in/out
flow tests should be performed.
8. If an expressive amount of medium is lost during air removal,
the medium must be completed to a final volume of 180 mL.
The o-ring must be completed dry with no medium residue to
avoid contamination.
9. For convenience, create a schematic map of the designed flow
orientation.
10. Cell adhesion can be affected according to the cell type and
material properties, such as roughness, porosity, and chemical
composition. Thus, in previous test trials, the ideal flow rate for
each specific purpose should be assessed.
11. The main difference in this protocol is that the system is set up
for a sinusoidal displacement approach until it detects a specific
132
Flavia Pedrini et al.